Cytometry Lab

Technical overview of the cell sorters


Sorter

FACStar

MoFlo

DiVa

Aria

Lasers in Use

1: 488nm*

2: 605nm**

1: 488nm

2: 633nm

3: UV 351-364nm***

1: 488nm*

2: 605nm**

3: UV 351-364nm

1: 488nm

2: 405nm

2: 375nm

3: 633nm

Nozzle Size

70 and 100 microns

70 microns

70 microns

70 and 100 microns

Sort Tubes

Clear Falcon tubes #2003 12x75mm Polystyrene

Opaque Falcon tubes #2005   12x75mm Polypropylene

Clear Falcon tubes #2003 12x75mm Polystyrene

Clear Falcon tubes #2003 12x75mm Polystyrene

Collection Tubes or Plates

12x75mm plastic or glass

1.5ml eppendorf

Plates: Terasaki or 6-96 well

12x75mm plastic or glass

1.5ml eppendorf

Plates: Terasaki or 6-96 well

12x75mm plastic or glass

1.5ml eppendorf

Plates: Terasaki or 6-96 well

12x75mm plastic or glass

15ml yellow cap

1.5ml eppendorf

Plates: Terasaki or 6-96 well

Optimum Cell Concentration

5x106 cells/ml

4x107 cells/ml

3x107 cells/ml

3x107 cells/ml

Typical Sort Rate

5,000 cells/sec

10-30,000 cells/sec

10-20,000 cells/sec

10-20,000 cells/sec

Number of Sort Populations

1 or 2 or Plates

1 to 4 or Plates

1 to 4 or Plates

1 to 4 or Plates

Sort Region Type

Rectangular

Polygonal

Polygonal

Polygonal

Running Pressure

7.5 - 12.5psi

55psi

30psi or 55psi

20psi, 35psi or 70psi

  • *     May be switched to 458, 476, 515nm.
  • **   May be adjusted continuously from 550-650nm.
  • *** May be switched to 405nm.
  • Allow 30 mins set-up time for any of these changes

For more details about possible choices of fluorochromes, see Fluorochrome Combinations.

Collection Tubes:

1-2mls of neat FCS or media containing FCS (whatever your cells will be happiest in).The tubes can be either glass or plastic (eg Falcon 2003 or 2005)

 Removal of dead cells and debris:

For smooth running of the sort and maximum yield of cells, run your cells through a dead-cell-removal column (glass Pasteur pipette with a plug of cotton wool in the end) or use a 100um filter prior to sorting.